En savoir plus !What is the problem when TGERR or T6ERR is briefly displayed at power up and all the calibration is lost?
On the HI 2110WI, HI 2151/30WC and the older HI 2151/20WC, when the Secure Memory Module (SMM) fails or is missing, you will receive a TGERR or T6ERR at power up.
The SMM part# for the HI 2110WI and HI 2151/30WC is 0519-0512-06.
The internal dip switch package S2-8 could be in the ON position. You can verify that setting by running Self-Test and observing DIP2 should be = 0
The SMM for the HI 2151/20WC is no longer available.